NOXBOX LTD: Device Recall

Recall #Z-0028-2026 · 09/09/2025

Class I: Dangerous

Recall Details

Recall Number
Z-0028-2026
Classification
Class I
Product Type
Device
Recalling Firm
NOXBOX LTD
Status
Ongoing
Date Initiated
09/09/2025
Location
Sittingbourne, United Kingdom
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1667 units

Reason for Recall

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

Product Description

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I