PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Device Recall

Recall #Z-0011-2026 · 09/03/2025

Class II: Risk

Recall Details

Recall Number
Z-0011-2026
Classification
Class II
Product Type
Device
Recalling Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Status
Ongoing
Date Initiated
09/03/2025
Location
Best, N/A, Netherlands
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 units (2 US, 6 OUS)

Reason for Recall

The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

Product Description

Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.